INSTALLATION
Note:
The manufacturer recommends that this procedure be performed by a qualified electrical
technician. The oven should remain disconnected from its power source throughout the procedure.
D
ISENGAGE THE
V
ACUUM
P
UMP
L
OCKS
The oven vacuum pump has four internal shipping locks on its base to cinch it to the floor of the
electronics cabinet. This prevents damage from impacts or shifting during transit. The locks must be
disengaged prior to placing the oven into operation. If the locks remain engaged, vibrations from the
vacuum pump can cause the oven to rattle.
Required Tools:
•
A Phillips Screwdriver
•
T-Handle Allen Wrench (Included in the roughing pump
document package)
1. Use the Phillips screwdriver to remove the front and back cabinet
access panels.
2. One at a time, insert the Allen wrench into each lock access port on
the roughing pump.
•
The 4 ports are indicated by labels below each port.
•
The order of disengaging the locks does not matter. Begin
with any of the ports.
3. Turn the Allen wrench to the right until each lock spins freely. This disengages
the locks. Disengage all 4 locks.
•
If a lock will not turn, move to another lock. Disengaging the
next lock may relieve tension on the first lock, allowing it to be
released.
4. Once all the locks are disengaged, reinstall the front and back cabinet access panels.
Note:
Refer to the vacuum pump
manufacturer’s manual if your pump
differs from the depiction below.