47
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
OPERATION
M
AXIMUM
O
BTAINABLE
V
ACUUM
The maximum obtainable vacuum is determined by the altitude of the oven workspace or laboratory
environment. The atmosphere is less dense at higher altitudes than at sea level. While a vacuum
pump will evacuate the same percentage of atmosphere from the oven chamber, less overall
pressure is expelled because of the reduced density.
Altitude
(Feet)
Altitude
(Meters)
Atmospheric
Pressure
Maximum
Vacuum Level
Attainable
Sea Level
Sea Level
14.70 psi
-29.9 inHg
1000ft
305m
14.16 psi
-28.9 inHg
2000ft
610m
13.66 psi
-27.8 inHg
3000ft
914m
13.16 psi
-26.8 inHg
4000ft
1219m
12.68 psi
-25.8 inHg
5000ft
1524m
12.22 psi
-24.9 inHg
6000ft
1829m
11.77 psi
-24.0 inHg
7000ft
2134m
11.33 psi
-23.1 inHg
8000ft
2438m
10.91 psi
-22.2 inHg
9000ft
2743m
10.50 psi
-21.4 inHg
10,000ft
3048m
10.10 psi
-20.6 inHg
*In gauge pressure
P
RESSURE
U
NITS
C
ONVERSION
C
HART
InHg
kPa
Kgf/cm
2
bar
psi
mmHG
mmH
2
O
1 inHg
1
3.3863
0.0345
0.3386
0.4911
25.400
345.32
1 kPa
0.2953
1
0.0102
0.01
0.1450
7.5006
101.97
1 Kgf/cm
2
28.9590
98.0665
1
0.9806
14.2233
735.55
10000.27
1 bar
29.5300
100
1.0197
1
14.5037
750.06
10197.44
1 psi
2.0360
6.8947
0.0703
0.0689
1
51.7150
703.09
1 mmHG
0.0394
1.3332
0.0014
0.0013
0.0193
1
13.5954
1 mmH
2
O
0.0028
0.0098
0.0001
0.0001
0.0014
0.0029
1
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......