12
5.1 Span Calibration
When the display shows “CAL”, press the key to enter Auto
calibration.
Please use , and , key to key in the mass
values which you are going to calibrate, then press key to confirm
the value. When the display is flashing and showing the weight values,
put the mass on scale. While scale is stable, the scale will return back to
weighing mode. The calibration procedure is finished.
NOTE:
●
Mass Value:
Any over (full calibration
×
10%)weight of key in
value is acceptable to calibrate.
●
Suggestion:
Generally. If the calibrated weight is over half
load capacity or full load capacity, it will be good for getting the
precise accuracy.
Another method for auto calibration, please see the section 6
“AUTO CALIBRATION”.
●
During in calibration procedure, you can press the
key
to exit calibration function.