ED
Service Manual
9
04/12/10
2.
Calibration
2.1.
General Calibration
Pressing and holding calibration switch press [POWER] key to go to calibration mode.
User can move to other mode by using [ZERO] key in the calibration mode.
User also moves to other sub-modes for each mode by using [TARE] key.
Please simply follow below procedure to move to other mode.
(1)
Calibration Mode: Pressing and holding “Calibration Switch” press [POWER] key.
(2)
It displays “CAL-0” after “CAL”, and it blinks the version of scale three times.
(3)
Selecting menu: press [TARE].
(4)
ENTER(Setting) : [TARE] key
MODE
Function
CAL 1
Display normalized AD
CAL 2
Display Keypad infomation-
CAL 3
Weight Setting Mode
“UnLoad” [TARE]
“MIDD” [TARE] after loading for 1/3 weight
“FULL” [TARE] after loading for Full weight
“MIDD” [TARE] after loading for 1/3 weight “END”
CAL 4
Option Setting ( Table 1
참조
)
CAL 5
Display filtered Raw AD
CAL 6
Function setting on each Key ( Table 2
참조
)
CAL 7
% Calibration
CAL 8
Battery calibration
CAL 9
Gravity constant
CAL 10
Set calibration factor
“Unit”
Æ
[TARE]
Æ
select 0, 1 (0:kg, 1: lb)
Æ
[TARE]
“CAPA”
Æ
[TARE]
Æ
select capacity
Æ
[TARE]
“MCAPA”
Æ
[TARE]
Æ
select mid-capacity
Æ
[TARE]
“W-dP”
Æ
[TARE]
Æ
Select Decimal Point
Æ
[TARE]
“ 1 d ”
Æ
[TARE]
Æ
Select division
Æ
[TARE]
“Dual”
Æ
[TARE]
Æ
Enable dual interval (0:disable, 1:enable)
Æ
TARE
CAL 11
Set nation(00 : OIML , 01 : NTEP , 02: KOREA)
Summary of Contents for ED Series
Page 1: ...ED Service Manual 1 04 12 10 ED SERVICE MANUAL ...
Page 3: ...ED Service Manual 3 04 12 10 8 Part List 25 ...
Page 7: ...ED Service Manual 7 04 12 10 1 4 Sealing Method ...
Page 8: ...ED Service Manual 8 04 12 10 ...
Page 15: ...ED Service Manual 15 04 12 10 3 The Schematics and Diagram 3 1 System Block Diagram ...
Page 18: ...ED Service Manual 18 04 12 10 4 Exploded View C A S C O LA B 2003 12 29 FO R A PPR O V A L ...
Page 21: ...ED Service Manual 21 04 12 10 z BOTTOM ...
Page 23: ...ED Service Manual 23 04 12 10 KEY PCB Part Location z DRAWING NUMBER 6150 PED 0000 0 KEY PCB ...