3.AD Test
Function : Load Cell Test
Used Key
Display Part
Description
: Upper Menu
Output value of the Calibrated
load cell is displayed.
Note 1. Check whether load cell output values are changed while loading and unloading
a weight on the load plate.
If the number is fixed or the number "0”is displayed, check again to note whether the
load cell is correctly connected
Note 2. When
key is pushed, the load cell output is displayed in the unit of mV/V
4.Communication Test
Function : Series Communication Test
Used Keys
Display Part
: Upper
Menu
Other Key : Test
Description
The value entered as Port No.1 is displayed in Communication Input 1
The value entered as Port No.2 is displayed in Communication Input 2
Simultaneously transmitted to Communication Outputs 1,2 upon key
inputting
Note 1. Execute this test in the state of executing the communication program(Hyper
Terminal) in the computer after connecting the series port of the computer and com
port on the back face of the indicator.
Note 2. Click‘1’to affirm whether the computer receives properly.
Note 3. Perform this test after designating the communication speed in advance
in Menu No.[2-2-04 or 2-2-09].
47
Summary of Contents for CI-600D Series
Page 1: ...2 ...
Page 36: ...5 2 How to Seal the Indicator Sealing 37 ...
Page 50: ...51 ...
Page 51: ...52 ...
Page 86: ...MEMO 87 ...
Page 87: ...MEMO 88 ...
Page 88: ...MEMO 89 ...
Page 89: ...MEMO 90 ...
Page 90: ...91 ...