18
INSTALLING
Loadcell connection
Loadcell connector connect with the loadcell port of indicator back
panel.
EX+
EX-
SIG-
SIG+
Load Cell
Connector
Resolution to loadcell output rate.
5V impression to loadcell
Max. loadcell output
Recommended
resolution
2 mV
1/2,000 (Max.)
5 mV
1/5,000 (Max.)
10 mV
1/10,000 (Max.)
Loadcell and connector connection
method
Reference : please refer to following loadcell color
because the wire color may be different by loadcell
production company and model.
Loadcell color table by company
Company
Connector
CAS
BONGSIN
JEANGWOO
KYOWA
INTERFACE
P.T
BLS
SHOWA
SHINKOH
TMI
TML
TFAC
HUNTLEIGH
No. 1
No. 2
No. 3
No. 4
No. 5
(EX+)
(EX-)
(SIG+)
(SIG-)
(SHIELD)
RED
WHITE
GREEN
BLUE
EXTERNAL
RED
WHITE
GREEN
BLUE
YELLOW
RED
WHITE
GREEN
BLUE
YELLOW
RED
BLACK
GREEN
WHITE
EXTERNAL
RED
BLACK
GREEN
WHITE
EXTERNAL
RED
BLACK
GREEN
WHITE
EXTERNAL
GREEN
BLACK
WHITE
RED
YELLOW
RED
BLUE
WHITE
BLACK
EXTERNAL
RED
WHITE
GREEN
WHITE
EXTERNAL
RED
WHITE
GREEN
BLUE
YELLOW
RED
BLACK
WHITE
GREEN
EXTERNAL
RED
BLUE
WHITE
BLACK
YELLOW
GREEN
BLACK
RED
WHITE
EXTERNAL