3.  Press the keypad the scale can not work: 

 
                                                                                                                                             

   

 
 

                                                                                                           

 

 

 

                                                                                                                                     

                                                                                                                                                                       

                                                                                                                                                             

 

                                                                                                                                             

 N 

 

 

 

                                                                                                         

                                                                                                                                                                       

Y

 

 

                                                           

                                                                                                                                               

 

 

 

 

 

 

 

 

 

 

The scale can not work   

when pressing the keys.

To short-circuit the 

keypad. 

The problem can be 

solved. 

The membrane 

keypad is damaged.

Change the 16 cores 

wire.  

The problem can be 

solved. 

The 16 cores wire 

is damaged. 

Change the CPU. 

The problem can be 

solved. 

The CPU is 

damaged. 

Change the 40 cores 

socket or RX2. 

Summary of Contents for EC-II

Page 1: ...ay data doesn t change or the internal resolution value is not in the natural range c The internal resolution data drift badly d The initial resolution is drifting 3 Trouble of power supply The phenom...

Page 2: ...ce it Check the output voltage of adapter is about 17V The AC jack for adapter is damaged replace it The adapter is damaged Please replace it Check if there is power output from the audion T5 Check if...

Page 3: ...age value between the pins of oscillator CR1 is 0 20v Change the oscillator Replace the CPU The main board is damaged Please replace it Displays well when power on The display is faulty Replace the 16...

Page 4: ...en pressing the keys To short circuit the keypad The problem can be solved The membrane keypad is damaged Change the 16 cores wire The problem can be solved The 16 cores wire is damaged Change the CPU...

Page 5: ...recalibrate The problem can be solved The load cell is damaged Check if the working environment is stable The voltage value is normal The battery voltage is less than 5 4V Recharge or replace the bat...

Page 6: ...e problem can be solved The 9 cors wire is damaged Change the display board The problem can be solved The problem is in display board Check if all the pins were welded bad Reweld them The problem is i...

Page 7: ...ain board and load cell is changed the scale must be recalibrated RS232 don t transmit any data Check the voltage of the 7 pins of the RS232 socket is around 10V The socket is damaged replace it Check...

Page 8: ...calibrated before WEIGHT UNIT WEIGHT COUNT 0 0000 00 200358 NET g Kg Ib HI OK LO NEGATIVE Local Remote COUNT 1 2 1If you want to change the calibration setting values please ENTER key 1 2 2If there i...

Page 9: ...n Count display window shows the total internal resolution which equals to original one and the increasing value Press SET key again to back to calibration mode 3 Change and choose the unit resolution...

Page 10: ...COUNT 3 2 2 Press the MOVE key to change the value kg Ib 3 2 3 Press the ENTER key to confirm and enter into next step Choose Resolution 3 3 Choose Resolution value 3 3 1 The display shows WEIGHT UNI...

Page 11: ...alue 1 2 5 3 4 3 Press the ENTER key to confirm and enter into next step Choose capacity 3 5 Choose capacity 3 5 1 The display shows WEIGHT UNIT WEIGHT COUNT HHHHH H CAP 200358 NET g Kg Ib HI OK LO NE...

Page 12: ...he ENTER key to confirm and enter into next step Enter the first calibration value 3 7 Enter the first calibration value 3 7 1 The display shows WEIGHT UNIT WEIGHT COUNT HHHHH H CAP1 200358 NET g Kg I...

Page 13: ...display window will show End Then press ENTER key to back to 1 2 of step1 without saving the second calibration value 3 9 Enter the third calibration value 3 9 1 The display shows WEIGHT UNIT WEIGHT...

Page 14: ...value Put on the weight refer to the display Press SMPL key for determination If SET key is pressed in step 3 9 press SMPL key calibration is finished and back to normal mode z Several seconds later t...

Page 15: ...r Local platform is g while the unit for Remote platform is kg 3 Put a weight on the pan same as what exactly shown in the UNIT WEIGHT window then press the ENTER key to confirm the operation The disp...