2 - ABOUT THE NEGATIVE AIR PRESSURE MACHINE/AIR PURIFIER

OptiClean

 uses highly efficient filters and a heavy duty, to remove contaminated air from the room. When ducting the unit, the resulting 

negative air pressure, or “vacuum effect,” helps limit the spread of air-based contaminants into surrounding areas. OptiClean

 is a 

portable solution designed to help convert normal hospital rooms into Airborne Infectious Isolation Rooms (AIIR).
If negative pressure is not required, such as in an open-air temporary hospital, business or home, the machine can be used as an air 

purifier pulling air in, removing many contaminants, and discharging cleaner air back into the room.

OptiClean

 Negative Air Pressure Machine features:

 

-

Long-life HEPA H13 (or H14 in option) filter according to EN 1822.

 

-

Standard M5 / Coarse 70% pre-filter according to EN 779/ ISO EN 16890

 

-

3 available units :1000, 1800 & 2500 m

3

/h (nominal)

 

-

Vertical design for smaller footprint

 

-

Portable and adaptable to nearly any location

 

-

Heavy duty locking casters for easy and smooth transport

 

-

Red lighted indicator to alert user when filters are overloaded

 

-

2,5 meters long power cord (access from behind the unit)

 

-

230V single phase - 50 / 60Hz

 

-

Galvanized steel, powder paint fully insulated cabinet

 

-

Options:

• Exhaust transition plate Ø250 mm for round ducting

• 

Diffuser

• UV Lamp

During normal use, this unit is intended to operate under the following site conditions:

 

-

Maximum altitude: 2000 m,

 

-

Minimum and maximum storage temperature: -20°C/+65°C,

 

-

Minimum and maximum operating temperature: 0°C/+40°C,

 

-

Indoor return air maximum humidity level: 27°C DB (dry bulb) at 65% RH (Relative Humidity),

 

-

Clean interior environment (no corrosive substances present).

The unit requires protection from any form of impact during use. 
This unit is designed to operate in an overvoltage category II and pollution degree 2 environment, in accordance with IEC 

standard 664. To ensure a pollution degree 2 environment, the unit must be protected from water and oil splashes, and the 

dust allowed to settle on it must be limited.

NOTE: the unit must be used indoor

 only. Keep away from sources of moisture, heat, flames.

This unit is not designed for Atex areas.

4

Summary of Contents for OptiClean 39UV

Page 1: ...ION OPERATION AND M A I N T E N A N C E I N S T R U C T I O N S Original document Negative Air Pressure Machine and Air Purifier OptiClean 39UV NOTE Read all instructions and data in this manual before operating ...

Page 2: ...PA Filter 7 5 6 Fuses replacement 8 6 TROUBLESHOOTING 9 7 SPECIFICATIONS 10 8 SPARES REFERENCES 11 The illustrations in this document are for illustrative purposes only and not part of any offer for sale or contract The manufacturer reserves the right to change the design at any time without notice Declaration of Conformity UE This unit complies with the provisions of European Directives 2006 42 E...

Page 3: ...ed by children WARNING ELECTRICAL SHOCK HAZARD Failure to follow this warning could result in personal injury or death Disconnect all electrical power to the unit before removing access panels to perform any maintenance WARNING PORTABLE UNIT AND METAL SHARP EDGES HAZARD Failure to follow this Warning could cause personal injury or death This unit must be used in a location where unsupervised child...

Page 4: ...Heavy duty locking casters for easy and smooth transport Red lighted indicator to alert user when filters are overloaded 2 5 meters long power cord access from behind the unit 230V single phase 50 60Hz Galvanized steel powder paint fully insulated cabinet Options Exhaust transition plate Ø250 mm for round ducting Diffuser UV Lamp During normal use this unit is intended to operate under the followi...

Page 5: ...ression clamp to attach Ø 250 mm round flexible ducting supplied by your local dealer to the transition plate c Run the flexible ducting as recommended by the mechanical plan and attach to a transition piece ceiling register box etc purchased or fabricated on site to complete a path for the air to exhaust from the negative air pressure area Caulk or tape all seams Run the flexible duct per applica...

Page 6: ...or sweep the floor around the unit to extend the life of the pre filter 1 Make sure the casters are locked 2 Locate the ON OFF switch on the rear side Fig 2 3 Rock the ON OFF switch up to the ON I position It should illuminate 4 Set the flow control on the desired airflow 5 Observe the unit running for a short time to make sure there are no unusual noises that air is freely flowing through the tra...

Page 7: ...rd from the unit and from the wall outlet 4 Listen to make sure the blower wheel has stopped spinning then remove the side door panel 5 Remove the filter holding frame on top of filters Fig 7 6 Remove the HEPA filter and immediately put it inside an approved containment bag 7 Remove the pre filter from cabinet and immediately put it inside an approved containment bag Dispose as ordinary industrial...

Page 8: ...ug These are fragile parts take care not to broke them 4 Gently remove the fuse holder 5 Replace the 2 fuses with new 4 A fuses see P10 for reference 6 Put back the fuse holder 7 Plug in the unit and rock the power switch up to the ON I position Fig 8 ON OFF switch C14 Cable plug Location of the fuse holder Fig 9 Removed fuse holder Fuses Gently push on the 2 small plastic clips on each side of th...

Page 9: ...it rear side if the unit does not turn on 2 Make sure the unit is plugged in to a working electrical outlet 3 Check for sufficient airflow Filters should be clean and unobstructed If you need to contact your authorized dealer for troubleshooting and or repairs be sure to have the model and serial numbers of your equipment available 9 ...

Page 10: ...LED Nominal flow rate 2500 m3 h 3 3 A 20mA filter LED Speed Variable Fig 10 Model Wiring Diagram Mains filter rated current 4A X Capacitor 1 stage X2 220nF Y Capacitor 1 stage Y2 1 5nF X Capacitor 2 stage X2 220nF Y Capacitor 2 stage Y2 1nF Resistor 1MOhms Setting Size 1000 400Pa Size 2500 500Pa BROWN G Y RED RED G Y BLUE Orange BLUE BLACK BROWN RED BLUE YELLOW EC radial module WHITE GREEN 10 ...

Page 11: ...te 1000 m3 h 1800 m3 h 2500 m3 h Pre filter M5 7588972 7588979 HEPA Filter H13 7588973 7588980 HEPA Filter H14 7588974 7588981 Transition plate 7588959 7590632 Diffuser 7588960 7590633 Fuses 7588986 Please contact your local Carrier organization to order 11 ...

Page 12: ...065 01 04 2021 Supersedes order No EN7592065 00 02 2021 Manufacturer reserves the right to change any product specifications without notice Manufacturer CARRIER SCS 01120 Montluel France Printed in the European Union ...

Reviews: