INST-PM-6046
Rev B 04/04/2018
Note: Photos may have a different appearance depending on chamber options.
1.
Turn off chamber and unplug power cord.
2.
Remove (3) screws in top access panel along front of chamber using Phillips
screwdriver bit as shown below.
Note: Retain screws.
3.
Remove top access panel by sliding towards front of chamber until (3) tabs on
each side are released as shown below.
4.
Loosen (2) clamps as shown below.
Note: Clamps may be colored different than
shown.
REPLACING CO2 IN-LINE FILTER
Disconnect electrical power before removing top access panel.
Screw Locations
Access Panel Tabs
Clamps