background image

 

 

 

 

INST-PM-6045 

Rev B   04/04/2018 

  

 

 

Note:  Photos may have a different appearance depending on chamber options.   

 

1.

 

Turn off chamber and unplug power cord. 

 

2.

 

Remove shelves from chamber.

 

3.

 

Remove shelf channels located near top of chamber if necessary.

 

4.

 

Remove false floor from chamber by pulling access port upwards while bending 
edges downward as shown below.

 

 

5.

 

Remove (4) screws securing intake diffuser using Phillips screwdriver bit as shown 
below.  

Note:  Retain screws.

 

  

 

 
 
 
 

6.

 

Remove intake diffuser by pulling towards front of chamber.

 

 

 

REPLACING ATOMIZERS 

Disconnect electrical power before removing interior parts. 

Intake Diffuser Mounting Screws (as viewed from inside chamber under intake diffuser) 

Intake Diffuser 

False Floor Access Port  

Summary of Contents for 6045

Page 1: ...MODEL 6045 INST PM 6045 ...

Page 2: ...Revision Log Revisions Description Letter Date A 09 05 2017 Initial release B 04 04 2018 New cover ...

Page 3: ... may be necessary Parts included in Preventative Maintenance Kit for Model 6045 are shown below Note Photos may have a different appearance depending on chamber options 1 Remove air filter cover by pulling outward on both sides near bottom as shown below INSTRUCTIONS FOR REPLACING AIR FILTER AND ATOMIZERS IN MODEL 6045 REPLACING AIR FILTER Air Filter Cover Air Filter Atomizers ...

Page 4: ...ed edge of air filter cover as shown below Note Orient cardboard side of air filter towards chamber 4 Install air filter cover with angled edge at top while aligning air filter with guide tracks on both sides and snap into place as shown below Note Right side is shown in photos Air Filter Guide Track Angled Edge on Air Filter Cover Air Filter Seam ...

Page 5: ... from chamber by pulling access port upwards while bending edges downward as shown below 5 Remove 4 screws securing intake diffuser using Phillips screwdriver bit as shown below Note Retain screws 6 Remove intake diffuser by pulling towards front of chamber REPLACING ATOMIZERS Disconnect electrical power before removing interior parts Intake Diffuser Mounting Screws as viewed from inside chamber u...

Page 6: ...d discard old atomizers using 2 8 crescent wrenches as shown below 10 Install new atomizers 11 Reinstall drip pan Note If necessary use flatblade screwdriver to install drip pan 12 Secure drip pan with retained 2 screws using Phillips screwdriver bit 13 Align tabs on intake diffuser with drip pan 14 Secure intake diffuser with retained 4 screws using Phillips screwdriver bit 15 Reinstall false flo...

Page 7: ...Rev B 04 04 2018 16 Reinstall shelf channels 17 Reinstall shelves 18 Continue to instructions for replacing additional accessory preventative maintenance items if applicable or plug in power cord and turn on chamber ...

Reviews:

Related manuals for 6045