Vue PACS MR Perfusion User Guide
Unrestricted Internal Use
AE9362 C
Uncontrolled unless otherwise indicated
Page: 36 of 43
Vue PACS MR Perfusion User Guide - AE9362.docx
Setting
Description
region.
Disable in cases when parts of the brain are mistakenly removed by the
ventricle-detection algorithm.
Clinical Settings
Setting
Description
Hematocrit
Set the hematocrit values.
The ratio between the large and small vessel hematocrit values affects the
scaling in the rCBV and rCBF maps.
Input State
Settings
Setting
Description
Vessel Settings
•
Number of AIF Candidates
—Set the number of automatically detected
AIF candidates.
•
Vessel's Signal
—Available options include
Seeds
and
Segmentation
.
Set the vessel's signal to
Segmentation
to automatically mark all
segmented arteries and veins on the image. Selecting this option hides
all AIF and VOF seed marks and enables the AIF and VOF
segmentation correction tools.
Segmentation
is the default selected
option in MR studies.
Set the Vessel's signal to
Seeds
to show the AIF and VOF candidates
and to manually add AIF and VOF points on the image.
Preprocessing Settings
•
Split data by slice time or slice number—Use
Slice Time
to arrange the
slices per time stamp. Use the
Slice Number
option to arrange the
slices per slice number rather than slice time (for example, in cases
when the original time stamp on the slices is missing or lacking
information). The assumed time interval between slices is 1.6 s.
Your choice is retained during your entire work session.
•
Starting Time
(used mainly in MR studies)—Insert the start time of the
study. The start time you insert here affects the start time displayed in
the graphic representation of the data.
•
Injection Time
—Insert the contrast material injection time to override
the injection time calculated automatically by the algorithm. The line
signifying the injection time in the graphic representation of the data is
repositioned based on your manually inserted time. Use the
Reset
button to revert to the original time calculated by the algorithm
automatically.
•
Perform Denoising
—Enable to preprocess images for noise reduction.
This setting is turned off by default for MR studies.