52 Chapter 5 Acquiring Images
7. Move the X-ray generator tube head to the patient and align it with the tooth of the patient and the
imaging plate.
8. Select the X-ray exposure time according to the region of interest and the patient type. Follow the user
instructions of your X-ray generator. The following tables provide
guidelines
for exposure times for an
X-ray generator at
70 kV
and
7 mA
. Add your values for the exposure time in seconds in the column on
the right.
Table 4
Exposure times
Parameter
DC-emitter, 7mA, tube length 20 cm
DC-emitter, 7mA, tube length 30 cm
recommended exposure time (s)
recommended exposure time (s)
Upper jaw
60kV
70kV
60kV
70kV
incisors
0.1
0.08
0.2
0.16
pre-molars
0.125
0.1
0.25
0.2
molars
0.16
0.125
0.32
0.25
Lower jaw
60kV
70kV
60kV
70kV
incisors
0.1
0.08
0.2
0.16
pre-molars
0.125
0.1
0.25
0.2
molars
0.125
0.1
0.25
0.2
Bitewing
0.16
0.125
0.32
0.25
Important: Make sure that the tube head is not shaking.
Important: These are suggested exposure times and need to be adjusted
for your specific X-ray generator. For dark images, reduce the exposure
time and for grainy images, increase the exposure time.
Summary of Contents for CS 7200
Page 1: ...User and Installation Guide CS 7200...
Page 6: ...2 Chapter 1 Conventions in this Guide...
Page 38: ...34 Chapter 3 Setting Up CS 7200...
Page 66: ...62 Chapter 5 Acquiring Images...
Page 68: ...64 Chapter 6 Maintenance...
Page 80: ...76 Chapter 8 Appendix...
Page 81: ...CS 7200 User Installation Guide 6M0940 _Ed04 77...
Page 82: ...78 Chapter 8 Appendix...
Page 84: ...80 Chapter 9 Contact Information...