CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed02
15
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity (IEC 60601-1-2)
The CS 7600 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CS 7600 should assure that it is used in such an
environment.
Immunity
Test
IEC
60601
Test
Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted
RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150 kHz
to 80
MHz
3 V/m
80 MHz
to 2.5
GHz
[V1]= 3 V
[E1]= 3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the CS 7600, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
d
(
3.5
V
1
-------)
P
=
d
(
3.5
E
1
-------)
P
80 MHz to 800 MHz
=
d
(
7
E
1
------)
P
800 MHz to 2.5 GHz
=