The calibration error for expiration and
inspiration at each flow rate are
displayed. The calibration error should
be less that 3%. If a greater error is
shown, repeat the procedure ensuring
that the syringe is emptied and filled in
a smooth manner without jerking the
handle. If an error greater than 3% is
still
shown,
inspect
the
turbine
transducer and clean if necessary.