VENTILATOR BATTERY CHARGE STATUS/LEVEL
The
External Power
indicator shows the level of external power while the ventilator is operating from an external power
source. When the ventilator is running form the internal battery, the External Power indicator is off. See below:
The
Charge Status
indicator shows the charge state of the internal battery. This LED is on when the ventilator is plugged
in to a power source and the internal battery is being charged. See below:
The
Battery Level
indicator shows the level of internal battery power while running from the internal battery. When the
ventilator is running from an external power source, the Battery Level indicator is off.
The length of time the ventilator
will work on internal power (up to 60 minutes) depends on settings, charge level and condition of the battery.
See below:
LED Color
Charge Status
Green
The external battery charge level is acceptable.
Amber
The external battery charge level is low.
LED Color
Charge Status
Flashing Amber
The ventilator is going through a pre-charge test of the
battery before starting the charge process.
Green
The internal battery is fully charged.
Amber
The internal battery is being charged but has not
reached a full charge level.
Red
The ventilator has found a charge internal battery fault
and cannot be charged.
LED Color
Battery Level
Approximate Battery
Time
Green
Internal battery level is acceptable
45 minutes
Amber
Internal battery level is low
10 minutes
Red
Internal battery level is critically
low
5 minutes