2.4.4
Fast Start
When
Fast Start mode
is enabled and a
pressure sensing disc is in use, the
instrument runs at an increased rate when
an infusion starts, taking up any slack in the
drive mechanism.
1. After accessing
System Config -
Syringe
options display, press
Fast
Start
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
• Option setting changes and
System
Config - Syringe
page
1
displays.
2.4.5
KVO
Enabling
KVO
allows some infusions to
automatically switch into
KVO
(keep vein
open) mode upon completion. The
KVO
option setting cannot be changed after the
instrument is powered on and a profile
selected.
NOTE:
In the following procedure, the
KVO
option is changed to
Enable
. Changing the
option setting to
Disable
, disables the
Change Rate
and
Change Volume
soft
keys, as displayed in step
1
.
1. After accessing
System Config -
Syringe
options display, press
KVO
soft
key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
CHECKOUT AND CONFIGURATION
2-5
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
S
y
stem Configuration - S
y
ringe
Fast Start mode
Enable
Disable
Allows use of the
fast start feature.
S
y
stem Configuration - S
y
ringe
KVO
Enable
Disable
Change
Rate
Change
Volume
Rate:
1
mL/h
Volume:
2%
Allows use of the KVO feature.
CONFIRM
S
y
stem Configuration - S
y
ringe
KVO
Enable
Disable
Change
Rate
Change
Volume
Rate:
1
mL/h
Volume:
2%
Allows use of the KVO feature.
CONFIRM
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
If
KVO
and
ALL
modes are both enabled,
the
ALL
mode does not deliver the entire
syringe contents, leaving a small
percentage of the syringe contents
available for
KVO
operation.
CAUTION
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......