TROUBLESHOOTING
6-8
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
354.6690
Logic Board
Pressure Sensor current
monitor: current is outside
acceptable range.
If error is repeatable: Replace
Logic Board or Pressure
Table 6-4.
Error Codes
(Continued)
Error Code
Subsystem
Explanation
Response
352.6700
Logic Board
Force Sensor
Force sensor circuit check
failure.
If error is repeatable: Replace
Logic Board or Force Sensor
356.6710
Digital Sensor
Digital sensor is outside
acceptable range during no-
power test.
If error is repeatable: Replace
digital sensor.
353.6650
Linear Position Sensor
Linear position sensor output
is zero or is outside
acceptable range.
If error is repeatable: Replace
linear position sensor.
355.6620
Syringe Sizer Sensor
Syringe sizer output is outside
acceptable range.
If error is repeatable: Replace
syringe sizer sensor.
354.6670
Pressure Sensing Disc
(Syringe Module only)
Pressure sensor output is
outside acceptable range.
Replace pressure sensing
disc.
Replacing the Force Sensor Assembly or Logic Board Assembly in a module with version
7
or earlier software
356.6780
PCA Handset
System detected PCA
Handset Key stuck at power-
up.
If error is repeatable: Replace
PCA Handset.
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......