SAD Day Light -
DL08127
3
The “SAD Light” has been designed and manufactured to conform to the highest standards of safety and performance
including electromagnetic compatibility (EMC).
The “SAD Light” conforms to the applicable requirements of the following:
This appliance meets the requirement of Medical Device Directive 93/42/EEC for medical appliances
2. Safety and performance standards
Emissions Test Standard
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Comply
Immunity Test standard
IEC 60601-1-2 Test Level
Compliance Level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Electrical fast transient/burst
IEC 61000-4-4
±2kV, for power supply lines
±1kV signal input/ output
100 kHz repetition frequency
±2kV, for power supply lines
Not applicable
100 kHz repetition frequency
Surge IEC 61000-4-5
± 0.5kV, ±1kV differential lines
± 0.5kV, ±1kV, ±1kV common mode
± 0.5kV, ±1kV differential lines
Not Applicable
Voltage dips, interruptions and
variations on power supply input lines
IEC 61000-4-11
0 % U
T
; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % U
T
; 1 cycle and 70 % U
T
;
25/30 cycles; Single phase: at 0°.
0 % U
T
; 250/300 cycle
0 % UT; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands be-
tween 0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands be-
tween 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Note:
UT is the a.c. mians voltage prior to application of the test level.