5. Storage  

 

It is advisable to store the mattress and pump in the product bag (accessory). Handle the 

packed product very carefully. Do not place any heavy objects on top of it. 

 

6. Maintenance 

 

1. Check the power cable and plug for damages and scratches. 

 

2. Check the mattress cover to make sure it is undamaged and that cover and cells are 

correctly assembled. 

 

3. Check that air is coming from the outlets on the side of the pump. 

 

4. Check for damaged or pinched tubes and connectors.  

 

Contact Care of Sweden or your local distributor for spare parts. 

 

 

7. Troubleshooting 

 

 

 

The pump does not start  

Check that the power cable is connected to an electrical 

outlet. Check that all fuses are OK. 

 

The user lies directly 

Pressure might be incorrectly set. Adjust it according to the 

on the bed  

weight table

 and wait a few minutes. Check again using the 

‘hand check’, see chapter 3.4. 

 

 

Check that the CPR-valve is closed. 

 

The mattress moves around   Check that the mattress is secured with the elastic bands to 

the underlying mattress. Or, if the mattress has an integrated 

foam core (

CuroCell S.A.M.

® CF16), secured by using 

the 

straps 

on the 

cover to fasten the mattress to the bed base.

 

 

The pump makes a noise,  

Check how the pump is hung on the bed.

 

Remove the pump 

or vibrates 

and listen whether this makes a difference. Can be resolved by 

putting

 

the pump on a flat, steady surface or by placing a 

towel between the pump and bed.

 

Pump bag (accessory) can 

also be used. 

 

If your questions are not answered by the above information, please contact 

Care of Sweden 

or your local distributor for

 further assistance. 

 

 

 

 

Problem 

Solution 

Summary of Contents for CuroCell S.A.M.

Page 1: ...Instructions for use 2020 12 08 CuroCell S A M Constant low pressure reactive air mattress system Instructions for use item number 95 001298...

Page 2: ...2 Product description 5 3 Installation and start up 7 4 Reuse cleaning and reconditioning 9 5 Storage 11 6 Maintenance 11 7 Troubleshooting 11 8 Technical description 12 9 Spare parts 14 10 Options an...

Page 3: ...OTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cable or plug is defective the pump casing is damaged or it is not functioning properly Cont...

Page 4: ...ful for all information involving accidents or incidents relating to our products by reporting them immediately to us at Care of Sweden 1 2 Intended use The system consists of a pump and a mattress an...

Page 5: ...can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to export careofsweden se 2 Product description 2 1 Pump 1 Front panel 2 Tub...

Page 6: ...air supply and pressure in the mattress increase When the knob is turned counter clockwise the air supply and pressure in the mattress decrease Appropriate air pressure is set between 1 and 8 The tabl...

Page 7: ...e bed or place it on a flat sturdy surface Place the power cable so there is no risk of stumbling over it run it over with the bed wheels or getting it jammed when raising or lowering the bed NOTE A p...

Page 8: ...nderlying mattress There should be about 1 2 fingers margin If you can feel the user s body resting on your hand you need to increase the comfort settings to next highest step Wait 10 minutes and perf...

Page 9: ...pose such as alcohol with or without tenside surfactants oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 If using a different agent choose one that does not affect the pump MATTRESS CO...

Page 10: ...utions chlorine max 1 or hydrogen peroxide max 1 5 Let it dry 5 Wipe off the work surfaces using a suitable detergent and or disinfectant 6 Put the mattress together If cells for any reason have be co...

Page 11: ...ectly Pressure might be incorrectly set Adjust it according to the on the bed weight table and wait a few minutes Check again using the hand check see chapter 3 4 Check that the CPR valve is closed Th...

Page 12: ...0x200 210x10 16 cm Weight 2 7 kg NOTE Care of Sweden reserves the right to modify the product specification at any time 8 2 Standarder Systemet har testats och godk nts enligt f ljande standarder r ti...

Page 13: ...70 C CE marked acc to Medical Device Regulation MDR 2017 745 Machine wash at 95 C Directly on bed base Tumble dry Place on top of existing mattress Do not iron Recommended user weight User informatio...

Page 14: ...31 105200 452 105200 CC 905616 105x210 cm 431 105210 452 105210 CC 905617 120x200 cm 431 120200 452 120200 CC 905618 120x210 cm 431 120210 452 120210 CC 905620 Air cells incl tubes with CPR Air cells...

Page 15: ...28047 90x210 cm CC 928067 90x210 cm CC 928048 105x200 cm CC 928068 105x200 cm CC 928050 105x210 cm CC 928070 105x210 cm CC 928049 120x200 cm CC 928069 120x200 cm CC 928051 120x210 cm CC 928071 120x210...

Page 16: ...and lasts longer During the expected lifespan of the product some parts of the pump will experience wear and will require servicing by a certified service technician in order to retain good functional...

Page 17: ...recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information contact Care of Sweden or you...

Page 18: ...18 13 Own notes...

Page 19: ...19...

Page 20: ...et www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Distrib...

Reviews: