Care of Sweden CuroCell Nova CX13 User Manual Download Page 1

 

 

User manual 

2015-08-10 

 

CuroCell

®

 Nova CX13 

CuroCell

®

 Nova CX19 

Alternating pressure system 

 
 

Item

 

number

 

user

 

manual:

 

95

001062

 
 

 

 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for CuroCell Nova CX13

Page 1: ...User manual 2015 08 10 CuroCell Nova CX13 CuroCell Nova CX19 Alternating pressure system Item number user manual 95 001062...

Page 2: ...roduction 4 Product description 5 Assembly Installation 7 Operation 8 Cleaning and reconditioning 19 Storage 12 Maintenance 12 Troubleshooting 13 Technical description 14 Spare parts 16 Options and ac...

Page 3: ...revents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective the pump casing is da...

Page 4: ...immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597...

Page 5: ...ces by sending your details to info careofsweden se 2 Product description 2 1 Pump and mattress 1 Mattress 2 CPR rapid deflation 3 Pump connector 4 Pump NOTE For certain patients such as amputees are...

Page 6: ...ilated cells they are located here 9 cells Foam core Tubing system CuroCell Nova CX13 CuroCell Nova CX19 with integrated foam core and tubing system 2 2 Pump 1 Connection tubes 3 Power cord 5 Suspensi...

Page 7: ...pressure must be increased at least one 1 step Lights when the mattress is ready for use Alarms for low pressure In order to solve it see chapter 8 troubleshooting 3 Assembly Installation When unpacki...

Page 8: ...click sound when the connectors are properly attached 4 Check that the switch on the pump is in the 0 position Off Connect the plug to an approved electrical socket 220 V 4 Operation Note Always read...

Page 9: ...h the user s pelvic region and make sure that the user does not lay directly on the underlying mattress There should be about 1 2 fingers margin If you can feel the user s body resting on your hand yo...

Page 10: ...the cell remain in its place in the mattress The air cell is activated by assembling the connection again 5 Cleaning and reconditioning It is important to follow the instructions below before using th...

Page 11: ...nstructions MATTRESS 1 Disconnect the tubes and let the air out also open the CPR valve 2 Remove the cover See instructions in chapter 5 1 for cleaning covers If the product is equipped with a foam co...

Page 12: ...he plug out of the electrical socket before replacing fuses 2 Remove the cover for the fuses with a screwdriver Pull out the fuse holder 3 Fit a new fuse of the same type T1AL 250 V and screw the prot...

Page 13: ...hand check see section 4 4 Mattress moves around Check that the air mattress is secured to the underlying mattress with the elastic straps If the mattress has a built in foam pad fasten it to the bed...

Page 14: ...ssification Class II Type BF IP2X MATTRESS SPECIFICATION Model CuroCell Nova CX13 Nova CX19 Dimension LxWxH 80 85 90 105 120x200 210x13 19 cm Weight 5 2 kg 80x200 cm 8 5 kg with built in foam pad NOTE...

Page 15: ...ser manual Machine wash at 70 C CE marked acc to MDD 93 42 EEC Machine wash at 95 C Directly on bed base Tumble dry Place on top of existing mattress Do not iron Recommended user weight Patient inform...

Page 16: ...cm CC 906013 Top part Olivia grey 85x210 cm CC 906014 Top part Olivia grey 90x200 cm CC 906015 Top part Olivia grey 90x210 cm CC 906016 Top part Olivia grey 105x200 cm CC 906018 Top part Olivia grey...

Page 17: ...ntilated cells Low Air Loss Micro ventilation in 9 of the cells 11 2 Accessories For CuroCell Nova CX13 Art no Description Size CC 9093 Sturdy transport bag 80 85 90 cm CC 9094 Sturdy transport bag 10...

Page 18: ...with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and perfo...

Page 19: ...a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information please contact Care of Swed...

Page 20: ...ernet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Supp...