6.9.5 Language
The texts can be displayed in different languages.
Figure 6 – 25: Language menu
6.9.6 Software Version
Select this option to view the installed software version.
Figure 6 – 26: Display of the installed software version
6.9.7 RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and confirm with
Select
.
Using the arrow keys, change the value and save the new
value with
Select
.
Figure 6 – 27: Setting the RPM limit values
Note
The limits selected in this menu only apply to the load range
between 6 and 150 watts. At higher loads the RPM limits auto-
matically adapt to the respective loads:
Load (watts) Green RPM range (1/min)
6 – 150
54 – 64 (adjustable)
151 – 250
61 – 62
251 – 350
71 – 72
351 – 450
81 – 82
451 – 550
91 – 92
551 – 650
101 – 102
651 – 750
111 – 112
751 – 850
121 – 122
851 – 950
> 130
951 – 999
> 135