xr100+
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The xr100+ ergometer is intended for use in the electromagnetic environment specified below. It is the responsibility
of the customer or user to ensure that the xr100+ ergometer is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level Electromagnetic Environment –
Guidance
Electrostatic discharge
(ESD) to EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
± 8 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/
burst to EN 61000-4-4
± 2 kV for power supply lines
± 1 kV for input and output lines
± 2 kV passed
Mains power should be that of
a typical commercial or hospital
environment.
Surge to EN 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV
N/A
Mains power should be that of
a typical commercial or hospital
environment.
Voltage dips, short inter-
ruptions and voltage
variations on power
supply input lines to
EN 61000-4-11
< 5 % UT
(> 95 % dip in UT) for 0.5 cycles
< 5 % UT
Mains power should be that of a
typical commercial or hospital envi-
ronment. If the user of the xr100+
ergometer requires continued opera-
tion during power mains interrup-
tions, it is recommended that the
xr100+ ergometer be powered from
an uninterruptible power supply or
a battery.
40 % UT
(60 % dip in UT) for 5 cycles
40 % UT
70 % UT
(30 % dip in UT) for 25 cycles
70 % UT
< 5 % UT
(> 95 % dip in UT) for 5 s
< 5 % UT
Power frequency
(50/60 Hz) magnetic
field to EN 61000-4-8
3 A/m
passed
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
The xr100+ ergometer has no
components susceptible to magnetic
fields.
NOTE: UT is the a.c. mains voltage prior to application of the test level.