WEIGHING THE PATIENT

1.  Wheel scale to patient's bedside.

2.  Remove stretcher "K"  from carrier "M"  and suspend it from stretcher support hooks "J".

3.  On the indicator "C", press the 

ON

 key on the keypad, then press the 

ZERO

 key.

4.  When indicator reads zero (0.0), remove stretcher from stretcher support hooks. The

indicator will show a negative weight equal to stretcher weight.

5.  With patient lying flat in bed, roll and place stretcher alongside patient.  Work stretcher

under patient by unrolling stretcher, much like changing a bed sheet.

6.  The patient's weight must be centered as nearly as possible along the length of the

stretcher, so that when lifting, the patient remains in a horizontal position.

7.  Roll scale so that base "A" is under bed, and stretcher support hoops "J"  are suspended

directly over patient. Set brakes on rear wheels "P".

8.  Widen base stance.  To adjust base stance, face rear of scale and pivot adjust handle "L"

towards the column.  Rotate handle unit until stance widens.  Return handle to a vertical
position to engage lock pin.  To return base to the narrow position, reverse above
procedure.

9.  Turn the cylinder "H" release valve counterclockwise to the open position.  When

stretcher support hooks "J" reach the mattress, return the release valve to the original
position.

10.  Engage the stretcher support hooks "J" in the four (4) holes of the stretcher braces "N".

11.  After closing the release valve, move the pump handle "G" up and down with slow strokes

of even pressure until the patient is about an inch above the mattress.  Make sure that no
part of the patient or stretcher is touching any part of the bed or mattress.

12.  When weight is stable, press the 

HOLD/RELEASE

 key to lock the displayed weight.

Note, that the weight displayed, will flash and remain locked until the 

HOLD/RELEASE

  

key is pressed again.

13.  Lower patient by 

slowly

 opening release valve counterclockwise.  When patient is back

on bed and stretcher support hooks "J" touch mattress, disengage hooks from stretcher
braces "N",  raise Boom "F", release brakes on rear wheels "P"  and remove scale from  
bedside.

14. Remove stretcher from under patient.  Replace stretcher in stretcher transport carrier "M",

and stretcher cup "R".

3

Caution: Open control valve 

SLOWLY 

to prevent sudden lowering of boom.

WARNING - This product is not intended for use in transporting patients!
It is to be used ONLY for weighing a patient. Failure to observe this warning may
result in serious injury to the patient and/or the Weighmobile operator. 

Summary of Contents for Detecto IB-400

Page 1: ...IB 400 Weighmobile Owner s Manual CARDINAL SCALE MFG CO 0046 M023 O1 Rev I PO BOX 151 WEBB CITY MO 64870 Printed in USA 05 00 PH 417 673 4631 FAX 417 673 5001 Web Site http www detectoscale com ...

Page 2: ...2 ...

Page 3: ...y with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules which are designed to provide reasonable protection against such interference when operated in a commercial environment Operation of this equipment in a residential area may cause interference in which case the user will be responsible to take whatever measures necessary to correct the interference You m...

Page 4: ...npack and remove all packing material 2 Check for damaged parts 3 Install batteries in indicator C Refer to the 758C Series Operation Manual 8555 M167 O1 for detailed instructions 4 Plug the coiled scale cable from the boom F into the 9 pin D shaped connector on the rear panel of the indicator Tighten the scale cable connector retaining screws to hold the cable securely to the connector on the ind...

Page 5: ...unterclockwise to the open position When stretcher support hooks J reach the mattress return the release valve to the original position 10 Engage the stretcher support hooks J in the four 4 holes of the stretcher braces N 11 After closing the release valve move the pump handle G up and down with slow strokes of even pressure until the patient is about an inch above the mattress Make sure that no p...

Page 6: ...COMPONENT IDENTIFICATION 4 C F G H P R N X J Figure No 1 T K A M ...

Page 7: ...COMPONENT IDENTIFICATION 5 X J F M C L P R A K Figure No 2 ...

Page 8: ... is stored in the stretcher transport carrier NOTE The lower eyebolts are secured by a special type of locknut and must NOT be tightened or adjusted Tightening will preload the weight sensor and result in inaccurate readings or possible permanent damage SERVICE INFORMATION The Detecto Weighmobile is designed and fabricated as a precision scale instrument It is comprised of four 4 major subassembli...

Page 9: ...PARTS LIST 7 Figure No 3 29 8 20 28 33 16 15 30 9 31 7 18 10 27 75 21 12 54 63 22 4 25 17 69 68 2 11 23 60 14 64 62 65 3 5 24 6 67 13 70 19 26 71 ...

Page 10: ...1 0046 C429 08 1 INNER BOOM 22 0046 C437 0A 1 STRETCHER SLEEVE WELDMENT 23 0046 C440 08 1 CAM BOX DOOR 24 0046 C447 08 1 LOAD CELL TUBE 25 0046 C465 0A 2 HOOP WELDMENT 26 0046 C471 08 1 758C BRACKET 27 0046 C467 0A 1 INDICATOR BRACKET WELDMENT 28 0046 D405 0A 1 CAM BOX WELDMENT 29 0046 D406 0A 1 LEFT LEG WELDMENT 30 0046 D407 0A 1 RIGHT LEG WELDMENT 31 0046 D408 0A 1 MAST WELDMENT 32 391RV204 2 1 ...

Page 11: ...EW 54 6022 0028 2 SPRING 55 6024 0009 3 3 8 FLAT WASHER 56 6024 0012 6 1 2 FLAT WASHER 57 6024 0017 2 5 8 FLAT WASHER 58 6028 0094 2 SPLIT RING 59 6028 0095 2 5 8 X 5 1 2 CLEVIS PIN 60 60S43 1 SPRING 61 650RV26 4 5 8 FLAT WASHER ZINC PLATED 62 6540 1070 2 FOAM HAND GRIP 63 6540 1078 4 1 1 4 SQUARE TUBING PLUG 64 6540 1080 1 HANDLE GRIP 65 6540 1119 1 HYDRAULIC CYLINDER 66 6540 1120 2 CAP PLUG 67 6...

Page 12: ...U N A M E D S I G U N E D F A C T R E D I N U S A U N A M E D S I G U N E D F A C T R E D I N U S A ...

Reviews: