2. Canister Full/Blockage - With Clamp on Tubing
If alert
continues,
go to step 3.
2
If alert
stays on, go
to step 5.
4
Make sure
clamp is
OPEN.
1
Wait
30 seconds
Wait
30 seconds
Look for kinks in
tubing. Unkink
if found.
3
Turn ON.
8
Clamp tubing and remove
canister
(see “Changing the
Canister” on front page)
.
6
Replace with new canister,
reattach tubing and open clamp
(see “Changing the Canister” on front page)
.
7
Turn OFF.
5
If you see:
If you see/hear:
Canister Full/Blockage
is fixed.
BLINK
BLINK
9
Wait
1 minute
BEEP
Call Cardinal Health
at 1.866.484.6798
SOLID LIGHT
+
+
WHAT YOU
SEE/HEAR
BLINK
BEEP
BLINK
+
+
Canister
Full/
Blockage
First, look at the canister to see if it is full. If the canister is full, change
the canister (see “Changing the Canister” on front page). If the canister
is not full, proceed with the steps below.
WHAT TO DO
NOTE:
A false Canister Full/Blockage alert can occur if drainage moves into the canister while the CATALYST™ is lying on
its front (the button side) or if it is turned upside down. If this occurs, you must change the canister to resume device
operation. Always keep the CATALYST™ upright.