6.1 Electromagnetic Compatibility
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ALLY TO GO™ is intended for use in the electromagnetic environment specified below.
The customer or the user of the ALLY TO GO™ should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The [ME EQUIPMENT or ME SYSTEM] uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
ALLY TO GO™
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ALLY TO GO™ is intended for use in the electromagnetic environment specified below.
The customer or the user of the ALLY TO GO™ should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 percent.
Electrical fast transient/burst
IEC 61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply lines
± 1kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
± 1kV line(s) to line(s)
± 2kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip in
U
T
) for 5 s
<5%
U
T
(>95% dip in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip in
U
T
) for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
ALLY TO GO™ requires continued operation during
power mains interruptions, it is recommended that
the ALLY TO GO™ be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note:
U
T
is the A.C. mains voltage prior to application of the test level.