TROUBLESHOOTING
6-2
Alaris
®
EtCO
2
Module (8300 Series)
Technical Service Manual
Table 6-1.
Technical Troubleshooting Guide
Follow the steps in the order they are listed until the problem/fault is corrected. Before making a final
diagnosis, visually inspect the instrument for damage and pay particular attention to the power cord
and plug. Following repair/replacement, use Maintenance Software to perform the required tests (see
Chapter
. Perform the Ground Current Leakage and Ground Leakage tests described in the PC Unit /
Pump Module Technical Service Manual to verify that the system is electrically safe.
Problem
Remedy
Display Problem
1.
Check cable connections.
2.
Replace Display.
3.
Replace Logic Board.
4.
Return to factory.
Key Stuck Alarm
1.
Turn instrument off and back on to see if problem clears.
2.
Replace Keypad Assembly.
3.
Replace Logic Board.
4.
Return to factory.
Instrument Malfunction
1.
Turn instrument off and back on to see if problem clears.
2.
Refer to alarm history, malfunction codes (Table 6-2), and subsystem codes (see
PC Unit / Pump Module (8000/8100 Series) Technical Service Manual).
Intermittent Operation
1.
Check connections to Logic Board.
2.
Replace Logic Board.
3.
Return to factory.
Summary of Contents for Alaris 8300 Series
Page 7: ...1 GENERAL INFORMATION...
Page 8: ......
Page 11: ...2 CHECKOUT CONFIGURATION...
Page 12: ......
Page 21: ...3 PREVENTIVE MAINTENANCE...
Page 22: ......
Page 25: ...4 PRINCIPLES OF OPERATION...
Page 26: ......
Page 29: ...5 CORRECTIVE MAINTENANCE...
Page 30: ......
Page 47: ...6 TROUBLESHOOTING...
Page 48: ......
Page 53: ...7 ILLUSTRATED PARTS BREAKDOWN...
Page 54: ......
Page 67: ......