8-6
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
CI Board Accessory Section
Reference the PC Unit and module-specific Sections of this
DFU
for the following references:
Alarms, Errors, Messages
Audio Characteristics
Definitions
Radio Frequency Note
The CI Board has self-testing capability, both at power on and
during normal operation. If communication with the CI Board
fails, a visual message displays to indicate a communication
error. A communication error to an outside access point does
not cause a Alaris
®
System alarm or visual message.
Alarms, Errors, Messages
NETWORK COMM ERROR
Communication error occurred
between CI Board and Alaris
®
Server.
Power Alaris
®
System off and then
back on. If error is repeated:
• To continue manual-only
operation, press
CONFIRM
soft
key.
• Replace PC Unit as soon as
possible.
• Refer instrument to be serviced
by qualified personnel, to:
- reprogram/reconfigure
CI Board
- verify wireless boot version
- replace CI Board, as
necessary
Message
Meaning
Response
Messages
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......