5-28
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
EtCO
2
Module Section
Audio Characteristics
Type
Sound
Notes
E
t
CO
2
Alarm (
HIGH PRIORITY
)
E
t
CO
2
Alarm (
LOW PRIORITY
)
E
t
CO
2
Error (Hardware Detected)
E
t
CO
2
Error (Software Detected)
A sequence of
5
beeps
One long beep approximately
every
4
seconds
A single alarm tone volume
Pairs of long beeps
Variable volume; can be silenced
for
2
minutes.
Variable volume; can be silenced
for
2
minutes.
Fixed maximum decibel volume;
cannot be silenced.
Fixed maximum decibel volume;
can be silenced for
2
minutes.
Alarms and Messages
(Continued)
CHANNEL ERROR
DISPOSABLE DISCONNECTED
HIGH ETCO2
HIGH FICO2
HIGH RR
LOW ETCO2
Hardware failure detected by
software.
Microstream
®
Disposable removed
from instrument during monitoring
mode.
E
t
CO
2
value is above specified
limit.
F
i
CO
2
value is above specified limit.
Respiratory rate is above specified
limit.
E
t
CO
2
value is below specified limit.
To silence alarm and continue
operation of unaffected instrument,
press
CONFIRM
soft key. Replace
with operational instrument, as
needed. Service by qualified
personnel required.
Attach Microstream
®
Disposable to
instrument.
Assess patient condition. Confirm
correct alarm limit values are
selected.
Assess patient condition. Confirm
correct alarm limit values are
selected.
Assess patient condition. Confirm
correct alarm limit values are
selected.
Assess patient condition. Confirm
correct alarm limit values are
selected.
High Priority Alarm
Meaning
Response
Alarms
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......