2-80
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Features and Displays
Features and Definitions
Reference the PC Unit Section of this
DFU
for system features and definitions.
Pump and Syringe Modules
Anesthesia Mode
When operating in Anesthesia Mode, a module can be paused
indefinitely without an alarm. Anesthesia Mode also makes it
possible to have additional drugs in each profile, which are only
accessible when operating in that mode.
Bolus Dose
Allows a bolus infusion to be programmed using either Drug
Library or drug calculation feature. It can be programmed with or
without a continuous infusion following a bolus.
Callback
A
callback
for a programmed delay (reference "Delay Options"
definition) can be scheduled to give an alert
Before
an infusion is
to be initiated,
After
an infusion is completed,
Before and After
an
infusion, or no alert (
None
).
Channel Labels
Available when Profiles feature is enabled. It provides a hospital-
defined list of labels, displayed in Channel (module) Message
Display, and identifies module with catheter location or other
helpful information.
Concentration Limits
Limits specified for range of concentrations allowed for a
particular drug in a profile.
Delay Options
Allows system to be programmed to delay start of an infusion
a)
for up to
120
minutes or
b)
for a specific time up to
23
hours
59
minutes.
Dose Checking
Always Dose Checking option causes a soft Guardrails
®
Alert
("Alert") to occur each time a dose limit is exceeded. Drug label
in Message Display provides an indicator ("
↑↑↑
" or "
LLL
") that
dose is beyond current Soft Limit.
Smart Dose Checking option causes an initial soft Alert to occur
when a dose limit is exceeded. Subsequent programming beyond
dose limit will not receive an Alert. Drug label in Message Display
provides an indicator ("
↑↑↑
" or "
LLL
") that dose is beyond current
Soft Limit.
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......