• The
Guardrails
®
Suite MX
incorporates dosing limits and
instrument configuration parameters based on
hospital/facility protocol. The software adds a
test of
reasonableness
to drug programming based on the limits
defined by the hospital/facility. Qualified personnel must
ensure the appropriateness of drug dosing limits, drug
compatibility, and instrument performance, as part of the
overall infusion.
Potential hazards
include drug
interactions, inaccurate delivery rates and pressure alarms,
and nuisance alarms.
• When loading a
data set
with the Guardrails
®
Suite MX,
ensure the correct profile
(for patient care area) is
selected prior to starting an infusion. Failure to use the
appropriate profile could cause serious consequences.
WARNINGS
2-76
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Guardrails
®
Suite MX
Warnings and Cautions
(Continued)
Epidural Administration
(Continued)
• The Alaris
®
System can be used for epidural administration
of
anesthetic and analgesic drugs
. This application is
only appropriate when using anesthetics and analgesics
labeled for continuous epidural administration and
catheters intended specifically for epidural use. Use only
an Alaris
®
Pump Module/Gemini Infusion System
administration set or syringe set, without a ‘
Y
’ connector or
injection port, for epidural infusions.
♦
Epidural administration of
anesthetic drugs
: Use
indwelling catheters specifically indicated for short-term
(
96
hours or less) anesthetic epidural drug delivery.
♦
Epidural administration of
analgesic drugs
: Use
indwelling catheters specifically indicated for either
short-term or long-term analgesic epidural drug delivery.
WARNINGS
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......