2-72
General Setup and Operation
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
1.
To stop infusion, press
PAUSE
key.
2.
Open plunger grippers and syringe barrel clamp.
• An audio prompt sounds (to silence, press
SILENCE
key).
• Red Alarm Status Indicator flashes.
•
CHECK SYRINGE
scrolls in Message Display.
SILENCE
OPTIONS
1
4
7
CLEAR
2
5
8
0
3
6
9
.
ENTER
CANCEL
SYSTEM
ON
3.
Remove syringe and separate administration set from
syringe.
4.
Reattach administration set to new syringe and load new
syringe (reference "Getting Started").
5.
Select syringe type and size (reference "Programming",
"Primary Infusion - With Guardrails
6.
Press
CONFIRM
soft key.
7.
Prime administration set (reference "Getting Started",
"Priming" procedure).
8.
Press
RESTORE
soft key.
OR
To enter
VTBI
and rate, press
RATE
soft key and use
numeric data entry keys, and then
VTBI
soft key and use
numeric data entry keys.
9.
To begin infusion, press
START
soft key.
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
System Start-Up / Setup
(Continued)
Changing Syringe During Infusion (
Syringe Module
)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......