2-68
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
1.
Ensure pressure sensing disc is installed correctly.
2.
Press
CHANNEL SELECT
key.
Installing a pressure sensing disc
after an infusion has started can
result in a bolus to the patient.
WARNING
Syringe Module
SILENCE
OPTIONS
1
4
7
CLEAR
2
5
8
0
3
6
9
.
ENTER
CANCEL
SYSTEM
ON
3.
Press
OPTIONS
key.
A
Current limit
550
_
mmHg
Pressure Limit Selection
CONFIRM
Change
value
Auto
Pressure
Current pressure
30
mmHg
Channel Options 1 of 2
Guardrails Drugs
Pressure Limit - 550 mmHg
A
>Select an Option or EXIT
EXIT
Multidose
Prime Set with S
y
ringe
Guardrails IV Fluids
PAGE
DOWN
4.
Press
Pressure Limit
soft key.
5.
To enter a new pressure limit value, press
Change Value
soft key.
OR
If Auto Pressure feature is enabled, press
Auto Pressure
soft key.
Pressure Sensing Disc Installed
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Selecting Pressure Limit
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......