Programming
2-69
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
NOTE:
If Auto Pressure is selected and current pressure is:
•
100
mmHg or less – system adds
30
mmHg to current
pressure, to create a new alarm limit
• greater than
100
mmHg – system adds
30%
to current
pressure, to create a new alarm limit
6.
Verify correct pressure limit input and press
CONFIRM
soft
key.
Selecting Pressure Limit
(Continued)
1.
Press
CHANNEL SELECT
key.
2.
Press
OPTIONS
key.
3.
Press
Pressure Limit
soft key.
Channel Options 1 of 2
A
>Select an Option or EXIT
EXIT
PAGE
DOWN
Guardrails Drugs
Pressure Limit - High
Multidose
Prime Set with S
y
ringe
Guardrails IV Fluids
4.
To select a pressure limit, press appropriate soft key.
5.
Press
CONFIRM
soft key.
A
Pressure Limit Selection
CONFIRM
Low
High
Med
Syringe Module
(Continued)
Pressure Sensing Disc Installed
(Continued)
Pressure Sensing Disc NOT Installed
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......