background image

DEFIB DSCHG 

Green indicator illuminates 
for two seconds when 
discharge is delivered from 
the AED. 

Product Description 

The Training Simulator is an easy to use training tool that allows practice in 
working with an AED and defibrillation procedures. 

 

NOTE:

 The Training Simulator is intending for training and sales demo purposes 

only.  Not for equipment testing or verification.  Not for human use.

 

 

Controls and Indicators 

 

 

The Training Simulator has eight waveform keys.  LED indicators in each key 
show you which waveform is running. 

 

 
 
 

 
 

 

 

 
Convert 

 

The convert feature allows the simulator to convert automatically from one 
shockable rhythm (the running rhythm) to another rhythm (the waiting rhythm) 
when a discharge is delivered from the AED.  To perform convert operation: 

1.  Press the key of a shockable rhythm. 
2.  Press the 

convert

 key.  Indicator of running rhythm pulses brighter. 

3.  Press key of the rhythm you want simulated immediately after defib 

discharge. 

4.  Deliver discharge from the AED. 

 
To cancel a convert operations before it’s completed, either press the convert 
key again or press the key of the running rhythm.  If a convert operation is started 
but not completed within two minutes, it will cancel automatically. 

 

Waveforms 

 

 

VF 

Ventricular Fibrillation, over 200 bpm. 

 

VT fast 

Ventricular Tachycardia (wide QRS).  Rate is approximately 185. 

  VT slow 

Ventricular Tachycardia (wide QRS). Rate is approximately 140. 

 

NSR 

Normal Sinus Rhythm.  Rate is approximately 72.  Non-shockable. 

 

ASYS 

 Asystole.  Non-shockable. 

 SVT slow 

Supra Ventricular Tachycardia.  Rate is approximately 170. 

 

artifact 

Unusable signal.  Rate is not applicable; frequency and/or amplitude 
exceed ECG parameters. 

  SVT fast

 

Supra Ventricular Tachycardia.  Rate is approximately 210.

 

Exercises 

 

 

The following examples demonstrate how the Training Simulator can be used for 
training. 
 
A. At Simulator: 

1.  Press 

VF

 key.  VF indicator glows steadily.  (VF is displayed on monitor.) 

2.  Press 

convert

 key.  VF indicator pulses brighter. 

3.  Press 

NSR

 key.  NSR indicator blinks on and off.  VF indicator glows 

steadily again. 

 
B. At AED in semi-automatic/AED mode: 

1.  Open lid of AED.  If packaged electrodes are under the lid, disconnect 

and remove them. 

2.  Connect simulator cable to the AED electrodes socket. 

The AED will give the following voice prompts: 

“Push flashing button to erase date and perform rescue.” 
“Do not touch patient.  Analyzing rhythm.” 
“Shock advised.  Charging.” 
“Stand clear.  Push flashing button to deliver shock.” 
 

C. With ECG monitoring cable (for Pro models only): 

1.  Open lid of AED.  If packaged electrodes are under the lid, disconnect 

and remove them. 

2.  Connect ECG cable to the AED electrodes socket. 
3.  Connect each of the 3 leads on the ECG cable to the appropriate 

terminal labeled RA, LA, LL. 

The AED will give the following voice prompts: 
 

“ECG Monitoring Mode.” 

4.  Press the 

NSR 

key.  NSR is now being simulated. View the ECG in the 

screen of the Pro AED. 

 

 

112-0050-002 Rev B 

 

ON-OFF 

Press to power on and to power off 
NSR is selected automatically when powered on.

 

LOW BATT 

Red indicator illuminates when 
battery voltage is low. 

convert 

Press before defib discharge 
to activate another rhythm 
immediately after defib 
discharge. 

3-LEAD TERMINALS:

 

LL, LA, RA 