Customer Calibration
17
MC16-EN– 1.04
8.5 Changing
the
Calibration,
Low Temperature Offset -
OFS.L
OFS.L
follows on from
CAL.L
when the
Page
key is pressed.
Press the
Up
or
Down arrow
key to display the current calibration low offset.
Press the
Up
or
Down arrow
key to alter the low offset value.
The value will then be stored without any further input.
8.6 Changing
the
Calibration, High Temperature –
CAL.H
CAL.H
follows on from
OFS.L
when the
Page
key is pressed.
Press the
Up
or
Down arrow
key to display the current calibration high temperature.
Press the
Up
or
Down arrow
key to change the temperature value.
The value will then be stored without any further input.
8.7 Changing
the
Calibration,
High Temperature Offset -
OFS.H
OFS.H
follows on from
CAL.H
when the
Page
key is pressed.
Press the
Up
or
Down arrow
key to display the current calibration high offset.
Press the
Up
or
Down arrow
key to alter the high offset value.
The value will then be stored without any further input.