3.0
Autotuning
All Carbolite Gero products are preconfigured for optimum performance when used
within their normal expected operating temperature range.
Depending on the intended application, it may be necessary for the product to be used
outside of this normal range, in which case some tuning may be required in order to
optimise the PID controls within the temperature controller e.g. if a high temperature
product is to be used at a significantly lower temperature than it was originally
designed for, or if a particularly heavy sample/load is to be heated.
To autotune the 132:
1. Enter the setpoint temperature that you wish to autotune the controller to e.g.
500°C.
2. From the home display, press and hold both the up and down arrow buttons
(more than 5 seconds) until
AUTOTUNING ACTIVE
appears.
3. When the autotuning process is complete, the controller will return to the default
home display.
Note:
To cancel the autotuning process, press and hold both the up and down arrow
buttons until the controller returns to the default home display.