background image

Programming 

MC18-GB – v1.04 

Press Scroll 

3

 until the display shows. 

TEXT

 

<

ENABLE/DISABLE SCROLLING TEXT

>

 

Use the Up 

S

Down 

T

 to select 

ON

 or 

OFF

 

 

4.6 Customer 

Calibration 

 
The 3216 series of controllers are calibrated for life at manufacture, there may however be 
sensor or other system errors, which affect the accuracy of the measured temperature.  
Customer calibration can be used to compensate for these errors. 
 

Dual Offset 

Dual point calibration uses two offset values at two corresponding temperatures; this 
changes the calibration linearly as the temperature increases or decreases.  
 
Press scroll 

3

until the display shows 

CAL.P

 (

Enter Calibration Code

) Use Up 

S

 

Down 

T

 

to enter the password code. (Calibration Pass Code = 95). When the correct password 

has been entered the display will show 

PNT.LO. 

If the wrong pass code is entered the display 

will revert to zero pass code, until the correct pass code is entered. 
 
When the correct pass code is entered and 

PNT.LO 

(

Adjust Low Point

) is displayed. Use 

Up 

S

 

Down 

T

 

to enter the Low Temperature Point, which you want to apply an Offset. 

 
Press scroll 

3

 until the display shows 

OFS.LO

 (

Adjust Low OFFset

). Use Up 

S

 

Down 

T

 

to 

enter the amount Offset you want to apply to the Low Temperature Point. 
 
Press scroll 

3

 until the display shows 

PNT.HI 

(

Adjust High Point

). Use Up 

S

 

Down 

T

 

to enter the High Temperature Point, which you want to apply an Offset. 
 
Press scroll 

3

 until the display shows 

OFS.HI

 (

Adjust High OFFset

). Use the Up 

S

 

Down 

T

 

to enter the amount Offset you want to apply the High Temperature Point. 

 
Once the calibration details have been entered, press scroll 

3

 until the display shows the 

next required parameter or return to the home list. The calibration data will now be protected 
by the pass code. To edit the data the above procedure must be followed. 

Summary of Contents for 3216 P1

Page 1: ...MC18 GB v1 04 Operating Instructions Temperature Controller Type 3216 P1 P5 English...

Page 2: ...2 MC18 GB v1 04...

Page 3: ...LER 8 4 1 MAXIMUM OUTPUT POWER SETTING 8 4 2 CUSTOMER ID 8 4 3 UNITS 8 4 4 LANGUAGE 8 4 5 SCROLLING TEXT 8 4 6 CUSTOMER CALIBRATION 9 4 7 HOLDBACK 10 5 PROGRAMMING 10 5 1 CREATING A PROGRAM 10 5 2 RUN...

Page 4: ...controller which uses PID algorithms to give excellent temperature control This controller can store and run a single program of 8 ramp dwell segments The 3216P1 can also be used as a simple temperatu...

Page 5: ...n alarm if activated Reset a program after the program has ended Arrow Keys S T The arrow keys are used individually to adjust the selected parameters and in combination to run a program Note If a par...

Page 6: ...The scrolling text can be interrupted at any time by a single press of any of the buttons but will not scroll again until the parameter is returned to In this manual the mnemonic will be shown first f...

Page 7: ...cted the display reverts to the home display See Section 2 1 Table showing parameters accessible in level 1 and Level 2 Operator LEVEL 1 Supervisor LEVEL 2 Home display Home Display Programming Progra...

Page 8: ...wer setting depends on the supply voltage refer to the product manual for details 4 2 Customer ID A furnace or oven identification number can be entered if required This maybe used to identify one of...

Page 9: ...he wrong pass code is entered the display will revert to zero pass code until the correct pass code is entered When the correct pass code is entered and PNT LO Adjust Low Point is displayed Use Up S D...

Page 10: ...then continue to control again The hold back will only apply once per Segment therefore when control has been re established the hold back will not apply again to that segment even if the furnace or o...

Page 11: ...play shows T SP 1 Target SP 1 Enter the temperature that you want the unit to ramp up to using Ramp rate 1 If Ramp Rate 1 has been set to OFF The unit will Heat Cool directly to the Target SP Dwell Ti...

Page 12: ...a program after it has completed Press and Hold S T for more than 1 second Or Press and quickly release Ack 3 Indicator RUN Off Scrolling Display None 5 3 Program Status While the program is running i...

Page 13: ...a power failure while the program is running and the power is subsequently restored the scrolling text will read POWER FAIL PROGRAM RESET To acknowledge this message and to reset the program Press ACK...

Page 14: ...ay shows T SP3 TARGET SP 3 Select 600 11 Press Scroll until the display shows DWEL 3 DWELL TIME 3 Select 60 12 Press Scroll until the display shows RMP 4 RAMP RATE 4 Select 3 13 Press Scroll until the...

Page 15: ...ed but may not be necessary The cable is usually 9 pin at the furnace end and 9 pin at the computer but other alternatives are shown in parentheses Furnace end of cable RS232 Cable furnace to PC Compu...

Page 16: ...The purpose of the 2 amp fuse is to guard against the connection of a mains supply to the panel plug The instrument configuration and parameters available to the operator depend on the customer requir...

Page 17: ...t sp1 A4 dwel 1 op hi rmp 8 LD AMP t sp8 LK AMP dwel 8 LD ALARM LD AMP If Configured LK ALARM ID HC ALARM ADDR ID UNITS CAL P PNT LO OFS LO PNT HI OFS HI Through to If Configured Through to Multi prog...

Page 18: ...Thermocouple The temperature measuring device used in the furnace or oven PID Proportional Integral Derivative the control system used by the controller Overtemperature O T The condition which a furn...

Page 19: ...Notes MC18 GB v1 04...

Page 20: ...hermal Engineering Services Telephone UK 0845 3308035 Int 44 1433 623335 Fax UK 0845 3308036 Int 44 1433 623336 Email service thermalserve com Carbolite Parsons Lane Hope Hope Valley S33 6RB England T...

Reviews: