OPERATING INSTRUCTIONS FOR THE 
CAPRICORN BENCH TOP CENTRIFUGE 
MODEL NO. CEP 2000

SETTING UP

WARNING - THIS EQUIPMENT MUST BE EARTHED

1.  Insert the mains lead into the IEC power inlet at the rear of the unit.
2.  Plug the unit into a grounded power socket - DO NOT SWITCH ON 

(OFF is indicated by ‘0’, ON is indicated by ‘1’).

3.  Ensure that the TIMER is in the OFF position, signified by the pointer 

being in a vertical position, and that the SPEED control knob is set to ‘0’ 

(see Figure1).

4. Switch the power on at the rear of the unit. At this point the green ‘POWER’ 

neon indicator will come on and, after a short delay, the amber ‘LID OPEN’ 
neon indicator will also come on, indicating that the lid may be opened.

LOADING

1.  Open the lid by undoing the catch on the front of the unit and lifting the lid.
2.  Place the sample tubes in the rotor, ensuring that:

(a) The correct tubes are inserted into their corresponding places in the rotor, 

if the rotor accommodates more than one size of tube.

(b) The rotor is loaded symmetrically, using tubes of equal size with contents 

of equal weight. If you are unable to balance the rotor, due to an odd number 
of tubes, fill an empty tube with water and use this as a balance. This water 
filled tube can, if necessary, be used continually for this purpose. 

(see Figure 2).

5

30

0

12

18

24

6

15

9

21

27

3

30

0

12

18

24

6

15

9

21

27

3

5

(Figure 1)

1

Summary of Contents for CEP 2000

Page 1: ...LANGUAGES PLEASE VISIT OUR WEBSITE CES INSTRUCTIONS SONT PR SENTES DANS D AUTRES LANGUES SUR NOTRE EMPLACEMENT DE WEB DIESE ANWEISUNGEN SIND IN ANDEREN SPRACHEN IN UNSERER WEB SITE VORHANDEN ESTAS IN...

Page 2: ...Knurled Nut Observation Window Speed Control Knob IEC Power Inlet Catch Neon Indicators Timer Knob CAPRICORN BENCH TOP CENTRIFUGE MODEL NO CEP 2000 SETTING UP PAGE 1 LOADING PAGE 1 STARTING PAGE 2 CO...

Page 3: ...n and after a short delay the amber LID OPEN neon indicator will also come on indicating that the lid may be opened LOADING 1 Open the lid by undoing the catch on the front of the unit and lifting the...

Page 4: ...4700 2200 6 5250 2750 7 5850 3500 8 6300 4000 9 6700 4500 10 7000 5000 2 Turn the TIMER knob in a clockwise direction to the required run time At this point the amber LID OPEN neon indicator will go o...

Page 5: ...lid may be opened 2 Remove the sample tubes and close the lid securing the catch on the front of the unit Once the lid has been closed the power should be turned off at the rear of the unit WARNING A...

Page 6: ...Centrifuge Model Number CEP 2000 Conforms to the following Product Specifications EMC EN61326 1 1997 Emissions Immunity EN55022 1998 Class B EN61000 3 2 1995 Class A EN61000 3 3 1995 Supplementary Inf...

Page 7: ...lasshouse Studios Fryern Court Road Fordingbridge Hants SP6 1NG UK Tel 44 0 1425 65 00 88 Fax 44 0 1425 65 00 92 or 0700 CAPRICORN 0700 2277426 e mail info capricorn labs co uk www capricorn labs co u...

Reviews: