MEASURE BACKGROUND
MAKE SURE NO SOURCES ARE NEARBY
AFTER MEASUREMENT--
PRESS SAVE OR PRINT TO ACCEPT
STAR
Figure 13-2
If you need to measure background, remove all sources from well (or drilled crystal) and
make sure no sources are nearby. Press the
F4 [START]
key.
The CAPTUS
®
600 will proceed to count the background for the counting time selected in
Background Count Time setup. During the count time, the live spectrum will be displayed.
The CAPTUS
®
600 will first count on the 400 keV scale and then on the 1600 keV scale. At
any time, you may abort the counting by pressing the
F2 [ABORT]
key.
When the background measurement is complete, Figure 13-3 will appear as long as the
counting rate has not exceeded the limits set in Background Trigger Level setup. If the
counting rate limits have been exceeded, Figure 13-4 will appear. This indicates that the
background is too high. Check for sources near the detector or contamination on the detector
and count again. If you are sure there are no sources nearby and the detector is not
contaminated, you may have to raise the background trigger level. The ROI used for
Background in comparing against the selected trigger level is the entire spectrum (all 256
channels).
B A C K G R O U N D T O O H IG H
B A C K G R O U N D : 25 .69 k c p m
T R IG G E R L E V E L : 9 9 9 9 c p m
P re ss A n y K ey to C o n tin u e
RANGE
COUNT
SAVE
EXIT
5
1600 keV
500 Cts FS 1024 keV
P
Figure 13-4
Figure 13-3
13-2
WIPE TEST
November 07
Summary of Contents for CAPTUS 600
Page 2: ......
Page 40: ......
Page 62: ......
Page 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Page 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Page 90: ......
Page 102: ......
Page 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Page 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Page 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Page 132: ......
Page 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Page 152: ......
Page 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Page 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Page 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Page 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Page 176: ......
Page 186: ......
Page 208: ......
Page 214: ......
Page 216: ......
Page 217: ......
Page 218: ......