Operation

!

!

Once  installed,  the  MB2-RSH  operates  in  exactly  the  same  way  as  a  conventional  MB2 
bioaerosol sampler.

!

!

For  operation  instructions,  please  refer  to  the  MicroBio  Manual,  document  number 
P0001W003.

!

!

!

!

Cleaning and bio-decontamination

!

!

The  control  unit  and  the  remote  sampling  head  may  both  be  cleaned  using  conventional 
sanitising wipes, such as Tuffie-5 or 70% IPA.

!

!

As the remote sampling head will normally be fixed within a compartment, it is suitable for 
use with Hydrogen Peroxide Vapour (H

2

O

vapour) bio-decontamination.

!

!

Tests  have  been  carried  out  by  Bioquell  (UK)  Ltd  on  sensitive  electronic  equipment  to 
determine effects of such processes.  The conditions of these tests were:

!

!

!

Gas Concentration

! !

1000 ppm

!

!

Conditioning time

! !

20 minutes

!

!

Gassing time

!!

!

30 minutes

! !

!

Aeration time

!!

!

150 minutes

!

!

As the MicroBio MB2-RSH has a fan within the remote head it is recommended that during 
the decontamination process the MicroBio MB2-RSH is set to sample 2000 litres several 
times  so  that  the  H

2

O

vapour  may  be  drawn  through  the  head,  fan  and  exhaust  for 

optimum decontamination.

!

!

Throughout  the  trials,  Bioquell  demonstrated  that  bio-decontamination  with  H

2

O

vapour 

does not appear to be detrimental in any way, effect operational aspects or aesthetics of 
sensitive electronic equipment.

!

!

As the control panel will not normally be exposed to bio-contaminants as the remote head, 
cleaning  using  sanitising  wipes  may  suffice,  but  it  is  also  suitable  for  use  within  H

2

O

vapour environments.

!

!

For further information on Bioquell H

2

O

vapour Material Compatibility please refer to their 

document  BDS-3-MATCOMP-V3.2  available  from  their  website 

www.bioquell.com

  or 

contact them on +44 (0)1264 835835.

Issue 2

!

/

                                                                                                                                                        12 12

Summary of Contents for MicroBio MB2-RSH

Page 1: ...ium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement MicroBio MB2 RSH Installation and Operation Supplement Document No P0002W006 Issue 2 Date 23 July 2010 Issue 2 1...

Page 2: ...Cantium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement Issue 2 2 12...

Page 3: ...n and Operation Supplement Document Issue 5 Technical Specification 6 Installation 6 Fitting the Connector for Unterminated Cables 6 Connecting the Control Unit and remote Sampling Head 10 Calibration...

Page 4: ...Cantium Scientific Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement Issue 2 4 12...

Page 5: ...2W006 MicroBio MB2 RSH Installation and Operation Supplement Document Issue Issue Description Date Revision by Approved 1 First release 20 June 2010 DCP SNJP 2 Added section on bio decontamination 23...

Page 6: ...e other with one of the connectors not fitted By not terminating one of the connectors enables the cable to be passed through glands or bulkheads as required by the end user Once the cable is routed t...

Page 7: ...ic Limited P0002W006 MicroBio MB2 RSH Installation and Operation Supplement 2 Fit connector housing and strain relief onto cable end 3 Remove cable clamp 4 FIt frame onto cable end and refit cable cla...

Page 8: ...o connector pins The connector body has pin numbers embossed onto the surface adjacent to the pins i Pin 1 Red ii Pin 2 Blue iii Pin 3 Screen 6 Fit pin housing back into frame noting that there are ra...

Page 9: ...icroBio MB2 RSH Installation and Operation Supplement 7 Slide the assembly strain relief and metal shell together 8 Align shell so that the brass thread of the frame body can be seen then replace the...

Page 10: ...is to the extension cable end labelled RSH The connectors have a screw locking mechanism Screw the two halves together by twisting the outer ring of the plug from the the remote sampling head connecto...

Page 11: ...nsion cable from the control unit and then plug the remote sampling head directly into the control unit as below If extension cables over 10m in length are used then it is recommended that the sampler...

Page 12: ...1000 ppm Conditioning time 20 minutes Gassing time 30 minutes Aeration time 150 minutes As the MicroBio MB2 RSH has a fan within the remote head it is recommended that during the decontamination proce...

Reviews: