– 21 –
4.3 Measurement in MANUAL Mode
6
Perform focusing
Turn the roller back and forth to position the three bright dots vertically between the
vertical alignment marks.
7
Measure the eye
After ensuring that the positioning
and focusing are as required, press the
START
switch.
[Continuous manual
measurement]
When the
START
switch is pressed
for about two seconds, the display on
the monitor will change from
“
MANUAL
” to “
MANUAL (C)
”, and
measurement will be performed until
10 measurements are obtained.
Perform alignment again as required
after the measurement is started.
Measurement can be cancelled by touching any switch other than roller or trackball
during continuous measurement.
PD: – –
VD:13.5
No. 00001
MANUAL (C)
RIGHT
SPH
CYL
AX
8:
-0.50
-0.75
36
9:
-0.50
-0.75
36
10:
-0.37
-0.75
37
[
-0.50
-0.75
36 ]
N
O
I
T
U
A
C
!
e
h
t
g
n
i
t
t
i
h
d
a
e
h
t
n
e
m
e
r
u
s
a
e
m
e
h
t
d
i
o
v
a
o
t
l
u
f
e
r
a
c
e
B
e
e
n
i
m
a
x
e
e
h
t
d
r
a
w
o
t
t
i
g
n
i
v
o
m
n
e
h
w
e
c
a
f
s
'
e
e
n
i
m
a
x
e
.
t
n
e
m
e
r
u
s
a
e
m
l
a
u
n
a
m
n
i
g
n
i
s
u
c
o
f
g
n
i
r
u
d
Alignment mark
Three bright dots must
be aligned vertically.
Alignment mark
PD: – –
VD:13.5
No. 00001
MANUAL
RIGHT
SPH
CYL
AX
1:
-0.37
-0.75
37
Summary of Contents for R-F10
Page 68: ... 56 This page has been left intentionally blank ...
Page 69: ...Blank Page ...