![Canon Omnera 500A Operator'S Manual Download Page 153](http://html.mh-extra.com/html/canon/omnera-500a/omnera-500a_operators-manual_61481153.webp)
Operator´s Manual
1000-095-071-EN Rev. 1.1
OMNERA®
2020-09-14
Page 145
5 Error Handling
5.1 Fault Handling
There are three types of NOTIFICATIONS - Shows the present occurence. For example; col-
lision. They are listed below in ranking order.
1. ERROR
Information in the OTC display (red). Sound: two beeps.
2. WARNING
Information in the OTC display (grey). Sound: one beep.
3. INFO
No information in the OTC display but can be found in the setting menu.
5.1.1 Notifications
5.1.1.1 - 1) Error
When an error occurs, an Error pop-up window will appear in the display.
60
˚
60
˚
cm
110
ms
20
mA
250
kV
50
Density
1
+
post mAs
5
8
mAs
AEC
DoB 1977-03-06
ID 987-65-4320
Hand AP
Age 36 Sex F
Acc No 987-65-4320
Jane Doe
SID
60
˚
60
˚
cm
110
ms
20
mA
250
kV
50
Density
1
+
1
1
post mAs
5
8
mAs
AEC
SID
ERROR
X
Checksum error
The {Node:Name} node has detected a
parameter problem
Restore back up parameters to the node
and restart system. If persistent error replace
the board
Fig. 5-1 Error pop-up window
The Error pop-up window will disappear when the user pushes the close button.
X