8.4 Wallstand
Column, Z stroke
1470 +40/-10 mm (non-tilt)
1400 +40/-10 mm (tilt)
1550 +40/-10 mm (tilt)
Rotation range detector holder wagon (On-
ly the tiltable detector holder wagon).
-20° - 90°
8.4.1 Attenuation Equivalent
detector holder
<=0.6 mm
8.4.2 Configuration
Wall stand
The Wall stand is the mechanical part of an
X-ray system.
8.4.3 Weight
Wall stand
Maximum 180 kg (160 +20/ -20 kg)
detector
Maximum 40 kg
8.4.4 Classification
Classification according to IEC/EN 60601-1.
Class
Class I equipment.
All dead metal parts of the equipment are
electrical connected to protective earth.
Applied part
Type B
Protection against ingress of water
IPX0
Mode of operation
Intermittent operation: 20% 1 min ON / 4 min
OFF
Use of anaesthetic mixtures
The equipment is not suitable for use in the
presence of flammable anaesthetics mix-
tures with air or with oxygen or with nitrous
oxide.
Classification according to IEC/EN 60601-1–2
Class
Class B