39
If the exposure dose is insufficient, artifacts may appear in the image. If a retake
is necessary, increase the tube current of the X-ray generator and capture the
image. If the tube current cannot be increased, increase the tube voltage and
capture the image.
Set the irradiation field as large as possible compared to the target diagnosis
domain. Otherwise, artifacts may appear at the edge of the irradiation field.
When using a grid with 34 lp/cm or 40 lp/cm resolution in Non Generator
Connection Mode, position the detector so that the two
marks face up and
to the left, and make sure that the direction of the grid lines is vertical, as shown
in the figure.
Important
Important
Important
Grid
Direction of the grid lines
mark
000_DIA-2019-019-E03_AX-C4343W.book Page 39 Friday, November 27, 2020 12:38 PM