CT-2000 Instruction Manual
Bath Operation
Caution:
Do not power on the CT-2000 without completing the installation
requirements. Make sure the
MAINS
voltage specified on the rear identification
label matches your MAINS voltage.
Note:
The CT-2000 stores the “last-used” temperature setting in memory. Upon
power-up, the bath will attempt to adjust to the last known/entered
temperature.
Cold Start
The Cold Start is the normal start-up mode for the CT-2000. During the Cold Start process, the CT-
2000 performs several diagnostics (see
). At the conclusion of a successful test
procedure, the results are briefly displayed on the LCD screen. The bath heaters are activated and
the Bath Unit begins controlling the temperature according to the most recent temperature entry.
The current bath temperature and the target bath temperature are visible on the liquid crystal
display (LCD) as shown in Figure 17.
Figure 17: LCD Temperature Setting
To Cold Start the CT-2000, toggle the power BATH switch up. The POWER lamp should light when
bath power has been activated. The five LED’s in the lower row of keys will blink on and then off,
and the instrument will emit a short “beep” tone.
Bath Lighting
To activate the Bath Unit fluorescent lights, toggle the LIGHTS switch up. The switch lamp indicator
lights up when power is supplied to the fluorescent lights.