8
| CAV 4.2 Operation Manual
Connect Hoses
Assemble the CAV 4.2 on a secure, level surface with sufficient room behind the instrument to maintain
a clear airflow and allow for unimpeded connection of the solvent inlet and waste outlet tubing.
See Figure 2.
Figure 2 — Tubing Connections
1.
Place the CAV 4.2 on a solid, level workbench. The workbench must be located near a MAINS AC
power connection capable of providing sufficient current (10 A at 115 V
AC
) to the CAV 4.2.
Depending on the materials tested, solvent used, and laboratory regulations, the work area may
also need to be equipped with a ventilation system.
2.
Place the waste receiver beneath the CAV 4.2. In order to promote proper drainage, the top of the
waste receiver must be lower than the waste outlets on the back of the instrument.
3.
Place the solvent bottles alongside or beneath the CAV 4.2.
Solvent inlets
Waste outlets
Exhaust ports