![Cannon CAV 4.1 Operation Manual Download Page 19](http://html1.mh-extra.com/html/cannon/cav-4-1/cav-4-1_operation-manual_3930099019.webp)
CANNON Instrument Company® | Getting Started
13
3.
Remove the front panel of the CAV 4.1 by gently pulling it away from the front of the instrument, as
shown in Figure 10. Set the front panel aside.
Figure 10 — Remove Front Panel and Set Aside
4.
Insert one end of the yellow tubing onto bath drain, leaving at least ¼" above the hose on the
nozzle. The hose must cover the large end of the nozzle, without making contact with the bath itself.
There must be adequate room above the hose to access the nozzle with the ¼"open end wrench
that opens the drain. Do not let the hose cover the entire drainage tube. Refer to Figure 11.
Figure 11 — Hose Inserted Properly onto Bath Drain (with Close Up)
Tube covers the end of the
nozzle
Room between hose and
bath for ¼" wrench