background image

New GentleYAG

 DCD Pressure Calibration Procedure  

Candela Corporation  

 

8503-01-0834, Revision 02 

Candela Corporation Proprietary 

 

Page 2 of 3 

 

DCD PRESSURE CALIBRATION 

1.  Connect the pressure transducer to the 300mv jack of the DVM.  Ensure the pressure 

transducer reads zero.  Attach the pressure transducer to the front bezel cryogen 
bulkhead.  

2.  Verify that a DCD Cryogen canister is installed. 

3.  Verify the pressure on the transducer is > 100psi before starting calibration. 

4.  Go to the Ckt Cal screen. 

5.  Press and hold the “DCD Press” button to start the calibration. 

6.  On the key pad, enter the DCD pressure reading from the external pressure transducer. 

7.  Press “Enter” at the bottom of the keypad to calibrate the DCD pressure. 

8.  Verify DCD canister is completely warmed up.  It must be between 115-125 PSI. 

Record/Verify on the TOGGLE screen DCD (PSI) reading matches the pressure 
transducer to within 

±

 1PSI.   

9.  Now remove the canister.  Let DCD pressure fully exhaust from laser and meter; about 

two minutes.  Record/Verify on the TOGGLE screen DCD (PSI) reading matches the 
pressure transducer to within 

±

 1PSI.  It must be between 0.0  - 1.0 PSI.  Re-install 

canister. 

10.  Exit Maintenance Mode. 

Reviews:

Related manuals for New GentleYAG