background image

Candela Corporation  

M

GentleLASE

 Focus Lens Replacement Procedure 

8503-01-0827, Revision A 

Candela Corporation Proprietary 

Page 5 of 5

J. 

Attach the plate (item 5), in orientation shown, onto the lens housing with 
hardware (2 ea.- items 6 and 7). Fully tighten the screws, then back out ½ 
turn each. 

K. 

Carefully rotate the plate (with housing attached), if needed, so that the plate’s 
(4) mounting holes are aligned with the (4) threaded holes in the receptacle 
body. Secure plate onto block with screws (4 ea.- item 6). Fully tighten 
screws. 

10. 

Install the receptacle onto the rail. 

11. 

Perform a Fiber Alignment Verification per the Service procedure 8503-01-0826. 

 

 

 

 

 

 

 

 