A. Specifications
Inputs
INPUT POWER – The Model 3102D is powered from a standard NIM Bin and power
supply, such as the Model 2100, 2000 or 1000.
INHIBIT – Logic low or ground inhibits the HV outputs; max logic low
≤
0.7 V; logic
high
≥
1.3 V or open circuit enables.
Outputs
HV OUTPUT – ±15 to ±2000 V dc, continuously adjustable; 1 mA output current ca-
pability; rear panel SHV connector.
÷
10 OUTPUT – ±1.5 to ±200 V dc, continuously adjustable; Z
out
= 20 M
Ω
; rear panel
SHV connector.
Controls
ON/OFF – Front panel toggle switch enables or disables output.
RESET – Pushbutton restores normal operation following a latched Inhibit and/or
overload fault condition.
VOLTAGE – Front panel 10-turn control permits continuous adjustment of the output
voltage.
POLARITY – Internal programming plug sets output polarity.
Indicators
HV OUTPUT – 3-digit panel meter; 0 to 2.00 kV.
POLARITY – Front panel LEDs indicate polarity continuously.
10
Specifications